Wrong Amount of DNA |
|
A basic mistake in quantification is to overestimate the amount of DNA
that is destined for sequencing. This will result in a fairly obvious
weak trace on the gel (signal strength values will be correspondingly
low), the major problem with this lies in the fact that the peaks will
be re-scaled once analysed and so the background noise associated with
the probe will also be re-scaled. This results in a characteristic
"noisy" sample. If you are using this sample for heterozygote
detection be very careful! Many false results may be seen.
|
![]() Figure: Insufficient DNA |
|
The peak heights should be nice and even throughout the length of the
run. If the peaks at the start of the sample are all off scale and
then the data suddenly drops down in strength, finally dying off only
300-400 bases into a run it probably means that the reaction has been
overloaded with template DNA (a so called 'ski-jump' profile when
looked at in the Full View image). When the sequencing reaction
becomes overloaded with template DNA the majority of the dye label is
incorporated within the first few rounds on the cycle sequencing. This
leads to the formation of many short extension products and usually
peaks that are off scale right at the start of the analysed
data. Simply reduce template quantities for sequencing.
|
![]() Figure: Too much DNA |
|
It must be noted however that this profile can also be seen when salt is present in a cycle sequencing reaction - this also leads to more
dye incorporation in the first few rounds of the cycling, although
usually it is not associated with the peaks that are off scale at the
start of the trace.
In general, it is best to quantitate your DNA on a gel by comparing the band with a band of a DNA marker of known quantity. If you are using a spectrophotometer try to use a salt buffer and not water to dilute your samples as this gives more consistent results! |